PTC255 & PTC255i

The Druck PTC255 series of Premium Temperature Calibrators are characterised by their unparalleled performance and outstanding ease of operation. By means of the intuitive menu structure, all necessary inputs can be made quickly and easily. The large touch screen has plenty of room to display the reference, target and devices under test temperatures. At the end of a calibration process, the PTC255 series provide the complete calibration certificate.

Description

Features

• Patented control technology: Fastest stabilisation times on the market – Time savings of up to 50 %
• Four functions in one calibrator (dry block /s calibration bath / infrared / surface)
• Large calibration volume / large calibration insert for simultaneous calibration of many devices under test
• Patented touch screen function for simple and convenient operation
• Automatic generation of the calibration certificate
• Device under test management with optional barcode scanner (P/N IOPTC-BAR-1)
• Integrated measuring model available (PTC255i)
• Automatic calibration with optional camera P/N IOPTC-CAM-2 and holder P/N IOPTC-CAM-1

Industries

Medical

Accurate pressure measurements improve diagnosis and treatment and can enhance patient safety through reduced risk of misdiagnosis. Accurate measurements also allow for better monitoring of conditions and optimize the use of medical devices such as venti…

Industrial

Overcoming challenges with precision and reliability Industrial operations rely on accurate and reliable pressure measurement to ensure safety, efficiency, and productivity. Druck, delivers high-performance pressure sensors and instruments that meet the …

Applications

FAQs

For recalibration of the Reference Probe, the Probe needs to be connected to the PTC 165 and calibrated as a complete system.
The errors in the Reference Probe are corrected for in the PTC when an adjustment is performed.

The Reference Probe is not calibrated as an individual item.

The adapter sleeve is available with 12 x 3.5 mm bores and 13 x 3.5 mm bores.

One end of the lead needs to have a Mini USB plug to go into the converter box, the other end of the lead needs to have a USB A connector suitable for connecting to the PC.

To install the Infra Red function a new setup of the three controller functions will be required.
All three functions will then need a new adjustment and new calibration certificates issuing.
The instrument will need to be returned to have the work listed below carried out:
a) Supply of Sleeve for Infra Red Temperature Measurement
b) Linearisation of Infra Red Function
c) Adjustment of LI, DB and IR Functions
d) Calibration of LI, DB and IR Functions

Downloads

File type
TypeTitleLanguageFormatDownload
DatasheetsDruck PTC255 255i Datasheet EnglishEnglishPDFDownload
User ManualsDruck PTC User Manual 168M9267EnglishPDFDownload
Quick start & safety guideDruck PTC Safety Manual Safety GuideMultilingualPDFDownload