Helium is not a suitable media to be used because over time (unable to predicate time
scale) it may affect performance of the PACE. The main problem with gases like helium
and hydrogen is the issue of migration through the isolation diaphragm. Helium is the
worst, but hydrogen is also prone to this because the molecules are very small and if
the pressure is high enough and the exposure time long enough then migration can
PACE5000 Ethernet Does Not Work
Check the software status as communication via the Ethernet was only supported from
Software Release 12 onwards. Therefore the Instrument Main Code DK367 needs to be at
V02.01.43 or higher. If the instrument has an earlier version of software installed then it
will require upgrading to the latest version.
At power on the PACE 5000 displays the error message “DPI.exe”. It no
longer powers up to the normal operation screen. This problem occurred
after attempting to upgrade the software.
The instrument may have failed to install the application. Try the following method to
see if the application will install.
Copy the attached Upgrade file to the USB stick.
Remove the cover from the PACE 5000. Insert the USB stick into the PACE 5000. Switch
on the instrument, then press and hold down the switch SW1 on the mainboard. Keep
the switch held down until the instrument displays Platform Upgrade, the switch can
then be released and the application should install
The calibration menu of some PACE 6000’s show an additional
value (highlighted in the image below).
The instruments seem to be
easier to calibrate with this value displayed.Can this feature be enabled
in the menu or is it only displayed for cert
It depends on the software version whether this feature is displayed or not.
When support for the CM3 was introduced at software version DK0388 V02.00.04 this
value was then displayed in the calibration menu.
The symbol shown is “sigma” which is the standard deviation, it gives an indication of
the stability of the pressure reading.
Initially when pressure is applied the value will be shown in black digits, then as the
pressure reading stabilises it will show in blue digits, then it will change to green digits
If you want to see this feature in the PACE 6000 calibration menu then you will need to
upgrade the PACE 6000 Software DK0388 to V02.00.04 or later.