The Foundation Fieldbus and Profibus DD Libraries are part of the instruments Main
Application (software DK0420). Upgrading to the latest version of DPI 620 Genii software
will ensure you have the latest DD Libraries.
The DD’s on the manufacturers website are not in the correct format for the DPI 620G
therefore cannot be directly added.
If there are any DD’s that are not in the current library then the you will need to provide
us with the DD for us to format, we can then supply the procedure to add it to the
library
Material Safety Datasheet (MSDS) DPI 620G-IS Battery Pack
The MSDS for the Battery Pack Assembly is attached > download here
DPI620G / DPI880 – Minimum Load for mV/TC Sourcing
We don’t have a specific value but would suggest that the impedance of the device
should be 10M ohms or greater.
The DPI 620/880 will provide the correct voltage at the output terminals with lower
impedance devices but then cable resistance will become more significant.
Generally voltage input devices tend to have a high input impedance so should not be
an issue.
DPI620G-IS not recognizing PM620 sensor
If the PM620 is working, but isn’t been recognized by the DPI620GENII-IS when connected
to the pressure base. Then this could be due to signal or 5Vdc missing from the
terminal. Check across the terminals for signals, as per shown below,
image.png
Power the DPI620GENII-IS on and you can then measure the 5V dc at the two outer
terminals shown in the diagram above. With instrument switched off you should see
between the two middle terminals resistance approximately 9.36K ohms.
If no voltage or resistance then it could be that either base or digital board that is faulty
and would need further investigation. Recommendation is to return the instrument back
to Groby Service Centre for repair
What is DPI620G Genii error «CH1 Source Error»
The error message is likely to be caused by a hardware issue.
The instrument will need to be returned for investigation/repair
Attachment contains the MSDS > download here
How to install / update the DPI620G HART Library
The HART Library needs to be installed on the instruments SD card and not the internal
memory.
There are two methods to upgrade the HART Library:-
1. Using the upgrade procedure in the User Manual K0541 Issue 2 – First ensure that the
instruments USB Client Port is set to Storage Mode. Access the USB CLIENT PORT setting
as shown below and ensure that the USB CLIENT PORT is set to Storage mode and not
Communications.
Connect the USB lead from the PC to the Mini Type B USB port on the instrument.
The Internal SD card should then be seen on the PC as a removable disk as shown in
the image below.
The HART Library zip file can then be extracted to the SD card.
2. An alternative method to upgrade the HART Library would be to remove the battery
from the instrument and then remove the SD card from its carrier within the battery
compartment. Plug the SD card into the PC or a card reader. Extract the HART Library zip
file to the PC and then copy the HCF folder to the SD card (Note – remove the existing
HCF folder first).
Refit the SD card to its carrier and then re-fit the battery to the instrument.
On power up the revision of HART Library should be shown in the Software Status
menu.
Material Safety Datasheet (MSDS) DPI610 and DPI620 Battery Pack
Attached are the 2 battery MSDSs depending on product.
DPI 610/615 = Sanyo
DPI620/Genii = Varta