The part numbers for the battery terminals with the pin pre-fitted are:
Positive Terminal IA2505-1-V0
Negative Terminal IA2505-2-V0
Cannot gain access to the calibration menu on DPI705 using the
calibration PIN 4321
First check the calibration enable link is fitted in the ticked position. The link is located on
the display side of the PCB.
If the link is fitted in the ticked position and the pin number 4321 still does not gain
access to the calibration menu then enter the number 7050 followed by pressing the
UNITS key
DPI705 displays incorrect number of decimal places and full-scale
calibration point is incorrect
The instrument range may have reverted back to its default value or has been set incorrectly. The range can be reset within the configuration menu as follows:-
First ensure the calibration enable link is in the ticked position. The calibration link is located on the display side of the PCB.
Press the FILTER, TARE and UNITS keys at the same time, «CAL» will be displayed, press the TARE key, using the FILTER and LEAK keys enter the pin 6971followed by pressing the UNITS key, «PDCR» will be displayed, press the TARE key, «RANGE» will be displayed, press the TARE key then use the FILTER key to change the range, when the correct range is displayed press the TARE key, press the UNITS key to exit to the pressure display.
The date on the factory Calibration Certificate is before Date of
Manufacture (DOM) of DPI705
The date stated on the calibration certificate is correct.
During manufacture the internal sensor of the DPI 705E is calibrated on a production
calibration rig.
The sensors have an inhouse calibration period of 15 months and if unused will be
recalibrated every 3 months.
This ensures that once fitted into a DPI 705E the calibration period will be at least 12
months when the instrument is shipped.
This is mentioned in section 3.1.4 of the user manual (see image below)
Note:- The number of Cal Due Days is displayed on the DPI 705E at power up
Retrieving data. Wait a few seconds and try to cut or copy again.
There shouldn’t be anything magnetically sensitive in the DPI705 or DPI705E to be used in this area.
When we do EMC testing we declare this so that we do not have to conduct the magnetic tests.