In most cases the Lithium battery used in our instruments will only be powering the Real
Time Clock to maintain the Date & Time. The Lithium battery will typically last for 10
years. If at some point after this the battery fails then the instrument will not keep the
time and date when switched off, it doesn’t mean that you have to replace the PCB.
For the DPI 611 the Real Time Clock is backed up by a rechargeable cell, when used on
batteries this is continuously charged until fully charged even with the instrument off
therefore the clock should retain its settings. If no batteries are fitted then the cell can
discharge, it will last approximately 30 days without batteries. If the instrument is left in
storage for a long time without batteries then the date/time may not be retained. It
takes approximately 50 hours to initially fully charge the cell.
The procedure below from section 1.13.1 of the DPI 611 user manual K0571 describes how
to recharge the clock battery.
Image_2020-01-24_13-43-14.png
The Scaling settings are missing in Software Rev19R for DPI611/612
There is a bug in release 19R of the DPI 611/612 software that affects the Scaling settings this will be corrected in the next release of DPI 611/612 software.
As as workaround until 20S is released you can downgrade the Application to Rev P (V02.08.05) – Released 25th March 2019
This can be downloaded from the download centre:- https://www.bakerhughesds.com/download-center
Unable to change Resolution on Software Rev R for DPI611/612
There is an issue with the Rev R software in that when the Leak Test Task is enabled the
Resolution is not displayed in the Settings menu.
The workaround is to turn off the Leak Test Task then the Resolution can be changed, if
need be the Leak Test Task can then be enabled again once the required resolution has
been set.
To turn off the Leak Test Task:
With the Pressure Channel displayed select the Screen Right icon to display the TASK
menu.
Select Custom Task -> Pressure -> Utility -> None -> Confirm selection by pressing the
tick/check icon twice.
Then access the Settings menu and set the Resolution to the required number of digits.
To access the Settings menu you need to touch the pressure reading, this will expand
the pressure reading window, you should then see a cog/wheel icon in the bottom right
corner of the display, touch this icon and Resolution should be listed in the menu.
Touch Resolution and you can then select the number of digits to be displayed.
DPI611.exe Error when Setting Time & Date on Software Rev R on DPI611/612
Upgrade the software to Rev S or later version if available.